Publication
IEEE Transactions on Magnetics
Paper

A Fast Scanning Magneto-Optic Kerr M-H Hysteresigraph for Thin Film Media

View publication

Abstract

This article describes an automated, scanning magnetometer that permits spatially resolved measurements of longitudinal magnetic properties to be made nondestructively on thin film disks used for data storage. The magnetization of a 2 mm diameter spot on a thin film disk is monitored by the magneto-optic Kerr effect while a magnetic field is swept in the plane of the film. The M-H response of a the magnetic film at a prescribed location can be measured in 1200 msec for a swept field amplitude of ±8 kOe. A PS/2 computer controls the swept field and R-θ position of the disk while simultaneously capturing and digitizing the Kerr response and the applied field strength. While the disk is translated from one location to the next, the PS/2 analyzes the hysteresis loop for the local coercivity, squareness, and coercivity squareness. The full hysteresis loops, in addition to these magnetic parameters, can be acquired for 100 locations in just under three minutes. © 1990 IEEE

Date

Publication

IEEE Transactions on Magnetics

Authors

Share